WESCOR NANODUCT
| Nanoduct™ Neonatal Sweat
Analysis System
|
| Sweat Induction Control |
Current profiled for patient comfort. For use with Pilogel Iontophoretic discs and with multiple fail-safe features to limit current even in failure mode. Nominal current is 0.5 mA for 2.5 minutes. |
|
CURRENT LIMIT (Failure Mode) |
Maximum failure mode current limited to 5 mA with fresh battery. |
|
SWEAT ANALYSIS |
Sweat Conductivity with readout expressed as mmol/L (equivalent NaCl). |
|
RANGE |
0 to 200 mmol/L (equivalent NaCl). |
|
INSTRUMENT LINEAR ERROR |
1% or less from 75 to 110 mmol/L (equivalent NaCl). |
|
INITIAL SWEATING RATE READING |
0.1 to 10 g/m2/min. |
|
CALIBRATION |
Single point automatic calibration at 80 mmol/L (equivalent NaCl). |
|
POWER SUPPLY |
One 9.0 volt Lithium battery (Alkaline battery can be used). 100 milliwatt power use during operation. Lithium batteries are good for approximately 300 tests. |
|
READOUT |
128 x 64 dot graphic display with multi-lingual support. |
|
AUDIO SIGNALS |
ALERT and ALARM audible signals. |
|
KEYBOARD |
ON, OFF, SELECT, and ENTER keys. |
|
CONNECTION |
5 pin connector to mate with induction/conductivity cell cable. |
|
OPERATING TEMPERATURE |
13 to 30 °C. |
|
STORAGE TEMPERATURE |
0 to 60 °C. |
|
REGULATORY |
Meets EN 61326 Standards for EMC compatibility. Meets IEC 60601 standards for safety. Manufactured under Wescor’s quality system (ISO 9001). |
|
PHYSICAL |
Housed in a portable carrying case with storage compartment for supplies and handle for ease of carrying. |
|
HEIGHT X WIDTH X DEPTH: |
16 cm x 23 cm x 33 cm. |
|
WEIGHT |
Less than 2.25 kg. |