Quadrasorb | Pharmaceutical Option | Specifications



QUADRASORB Sl Specifications

Analysis Specifications
Transducer Accuracy: 0.11% full scale (1,000 torr transducer) Four Stations/One Manifold
0.15% reading (10 torr transducer - optional)
Pressure Resolution: 0.015 torr (1000 torr range)
0.00015 torr (optional 10 torr range)
Sensitivity: <1 X 10-7 moles adsorbed/desorbed gas
Ultimate Vacuum: 1x10-3 torr achieve by dedicated 2 stage rotary, direct drive pump or 1x10-9 torr achieve by oil-free turbomolecular vacuum pump in QUADRASORB SI - KR/MP
Adsorbates: Nitrogen and any other non-corrosive gas with appropriate coolant
Surface Area Range: <0.05m2/g with N2 (no known upper limit)
Pore Volume: Detectable limit less than 0.0001 cc/g
Pore Size: Diameter range 3.5 - 4000A / 0.35 - 400 nm (with nitrogen)
Coolant Level: Automatic compensation using patented TempComp™ method
Minimum P/P0 1 x 10-3 QUADRASORB SI
4 x 10-5 QUADRASORB SI-KR/MP
* Pressure transducer and vacuum pump specifications from their respective manufacturers.
Physical Specifications
Electrical
100 - 240 VAC
50/60 Hz
Dimensions
Height: 73.4 cm (28.9 in)
Open Height: 111.6 cm (44.0 in)
Width: 63.7 cm (28.1 in)

Depth: 47.7 cm (18.8 in)
Open Depth: 66.5 cm (26.2 in)

Weight
45.3 kg (100 lbs.)
Operating Environment
10 - 38°C operating range at 90% maximum relative humidity

Giangarlo Scientific | Analytical | Environmental | Life Science | Process | Safety | Wastewater | Service | Contact Us