PoreMaster Series | Specifications | Overview | Features/Benefits


Pore Size Analyzers


PoreMaster
GT Specifications

 

Transducer
Pressure Range

PoreMaster 60 GT

PoreMaster 33 GT
0 - 50 psia*
     § Accuracy:
      Resolution:
950 - 4.26 µm dia.
± 0.11 % fso**
0.000763 psia
950 - 4.26 µm dia.
± 0.11 % fso
0.000763 psia
0 - 1500 psia
     § Accuracy:
      Resolution:
10.66 - 0.142 µm dia.
± 0.11 % fso
0.0229 psia
10.66 - 0.142 µm dia.(Upgrade)
± 0.11 % fso
0.0229 psia
0 - Maximum psia
     § Non-linearity:
       § Hysteresis:
          Resolution:
10.66 - 0.0036 µm dia.
± 0.05 % fso
± 0.10 % fso
0.916 psia
10.66 - 0.0064 µm dia.
± 0.05 % fso
± 0.10 % fso
0.511 psia
*Pounds per Square Inch Absolute ** Full Scale Output § Manufacturer's Specifications
Sample Cells
Sample Volume Intrusion Capacity Volume Resolution
3.2 cc 2.0 cc 9x10-5 cc
3.2 cc 0.5 cc 3x10-5 cc
6.6 cc 2.0 cc 9x10-5 cc
6.6 cc 0.5 cc 3x10-5 cc
Physical
Width: 20.25 in. 51.44 cm
Depth: 25.50 in. 64.77 cm
Height: 36.00 in. 91.44 cm
  49.00 in. (lift extended) 124.46 cm (lift extended)
Weight: 398 lbs. 179 kg
Electrical
Mains: 100 - 240 volts AC, 50 Hz/60 Hz
Power: 1000 VA
Required
 

Compressed dry gas, cryogen, PC with Windows 95/ 98/2000/XP, 32MB of RAM, 3 GB hard disk space, 3 ½" disk drive, serial and parallel ports, printer .


Giangarlo Scientific | Analytical | Environmental | Life Science | Process | Safety | Wastewater | Service | Contact Us