| The Monosorb, a direct reading dynamic flow
surface area analyzer, utilizes a modified B.E.T.
equation for extremely rapid single point determinations
of surface area. Features:
- Speed
The Monosorb:
*requires as little as 6 minutes to complete an
analysis.
*uses small quantities of sample resulting in
short outgassing and analyses times.
*utilizes the most modern and reliable method for
extremely rapid and accurate B.E.T. surface area
measurements.
*doesn't require void volume measurements and
ideal gas corrections.
|

- Versatility
The
Monosorb permits the user:
*
to choose the correct concentration
of adsorbate to give results identical to the
multipoint method.
*to use many different adsorbates
in addition to nitrogen, such as, argon,
krypton, carbon monoxide, carbon dioxide and
other non-corrosive gases.
*to
use a wider variety of sample cells allowing
measurements to be made on many diverse
sample.
- Automation
*with
the push of just one button, the sequence of
adsorption, warming of the sample cell,
desorption and display of the surface area,
all occur without operator intervention.
*AUTOCALIBRATE feature
guarantees proper calibration at all times.
*built-in Flow Controller
provides the ultimate in flow stability.
The high
degree of sensitivity of the Monosorb allows the use
of samples weighing less than one gram and provides
the measurement of surface areas as low as 0.1 square
meters in the sample cell regardless of the adsorbate
gas used. The upper range of the Monosorb is over 250
square meters in the cell. The reproducibility is
better than 0.5%. The accuracy of the Monosorb is
limited by the applicability of the single point
method, which is usually better than 5% and on most
samples errors less than 1-2% are found.
Specifications
Power
- 100-120V or 220V, 50/60 Hz
Dimensions -
25½"H x 12"D x 24"W (65cm x 31cm x 61
cm)
Weight - 53 lbs.
(24 kg)
|