| Specification |
Autosorb-6B |
Autosorb-6B-Kr |
PRESSURE (Specifications by
transducer manufacturer ) |
Range:
0 - 1,000 torr
|
Range:
0 - 10 torr
|
| |
Accuracy |
0.001
x full scale |
0.0015
x reading |
| |
Linearity:
0.05% full scale |
0.0005
x full scale |
0.0015
x reading |
| |
Minimum
resolvable pressure: |
0.015
torr |
0.0002
torr |
| |
Minimum
resolvable
relative
pressure (P/Po): |
0.00002
nitrogen |
2.6
x10-7nitrogen
7.6 x 10-5
krypton |
| ULTIMATE VACUUM: |
Less
than 0.001 torr |
Less
than 10-5 torr |
|
| SORBATE
GASES: |
Nitrogen
or any non-corrosive gas with appropriate coolant |
| SURFACE AREA RANGES: |
Less
than 0.0005 m2/g with Kr to no known
upper limit
Less than 0.05
m2/g with N2 to no known
upper limit |
| PORE ANALYSIS: |
Less
than 0.0001 cc/g detectable limit
3.5 - 4000Å
pore diameter range with nitrogen |
|
| ACCESSORIES: |
|
| Required: |
Adsorbate
and helium (minimum purity 99.999%), coolant
Windows®
compatible personal computer |
| Optional: |
Autosorb
DEGASSER, uninterruptable power supply,
20 inches
(50cm) high caster mounted cabinet |
|
| Physical |
Autosorb-6B |
DEGASSER |
| DIMENSIONS:
|
|
|
Width
|
25.5
inches (64.8 centimeters) |
30.0
inches (76.2 centimeters) |
Height
|
45.5
inches (115.6 centimeters) |
29.0
inches (73.7 centimeters) |
Depth
|
30.0
inches (76.2 centimeters) |
17.6
inches (44.7 centimeters) |
| WEIGHT: |
380
pounds (172.7 kilograms) |
199
pounds (90.5 kilograms) |
| ELECTRICAL: |
100-240
volts, 50/60 Hertz |
100-240
volts, 50/60 Hertz |
|
| TEMPERATURE: |
10 -
38°C. operating range at |
MANTLE:
420°C. maximum |
|
|
90%
maximum relative humidity |
SELECTION:
steps of 1°C. |
|
|
|
RAMP