HI 9835 Waterproof,
Autoranging, EC/TDS/NaCl/ºC Meter
![]() |
|
The Hanna HI 9835 can measure conductivity, total dissolved solids, NaCl, and temperature, all with one probe. The temperature of all conductivity and TDS measurements is automatically or manually compensated for by the meter's ATC circuitry. The temperature coefficient is user selectable from 0.00 to 6.00 %/°C (for EC and TDS only) and the TDS factor can be set from 0.40 to 0.80.
The meter is constructed with a waterproof, rugged housing which allows the versatility required for both laboratory and field use. Moreover, HI 9835 is equipped with a large LCD that helps the user through operation and calibration with large comprehensive symbols. Besides displaying the measurements, the LCD simultaneously displays the temperature. The display gives a low battery warning and switches itself off after 5 minutes of non-use to save the batteries.
HI 9835 is supplied complete with HI 76309 Conductivity/TDS probe with 3.3' (1 m) screened cable, AA size Alkaline Batteries (4 pcs) and instruction manual in a rugged carrying case.
| Specifications: | HI 9835 | |||
| EC (Autoranging) |
TDS (Autoranging) |
NaCl | Temperature | |
| Range | 00.0 to 29.99 µS/cm 30.0 to 299.9 µS/cm 300 to 2999 µS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm Up to 500.0 mS/cm actual (*) Conductivity |
0.00 to 14.99 ppm 15.0 to 149.9 ppm 150 to 1499 ppm 1.50 to 14.99 g/L 15.0 to 100.0 g/L Up to 400.0 g/L actual (*) TDS (with 0.80 factor) |
0.0 to 400.0 % | 0.0 to 60.0°C |
| Resolution | 0.01 µS/cm (from 0.00 to
29.99 µS/cm) 0.1 µS/cm (from 30.0 to 299.9 µS/cm) 1 µS/cm (from 300 to 2999 µS/cm) 0.01 mS/cm (from 3.00 to 29.99 mS/cm) 0.1 mS/cm (over 30.0 mS/cm) |
0.01 ppm (from 0.00 to
14.99 ppm) 0.1 ppm (from 15.0 to 149.9 ppm) 1 ppm (from 150 to 1499 ppm) 0.01 g/L (from 1.50 to 14.99 g/L) 0.1 g/L (over 15.0 g/L) |
0.1 % | 0.1°C |
| Accuracy (@ 20°C/68°F) |
±1% of reading ± (0.05 µS/cm or 1 digit, whichever greater) |
±1% of reading ± (0.03 ppm or 1 digit, whichever greater) |
±1% of reading | ±0.4°C |
| Typical EMC Deviation | ±1% of reading | ±1% of reading | ±1% of reading | ±0.1 °C |
| Calibration | 1 point with 6 memorized
buffers: 84, 1413 ,5000,12880, 80000 or 111800 µS/cm |
1 point with HI 7037 buffer (optional) |
2 points at 0 and 50°C (plus ±1°C adjustment) |
|
| Temp. Compensation | Automatic or Manual from 0 to 60°C (can be disabled to measure actual conductivity) | |||
| Temp. Coefficient | 0.00 to 6.00 %/°C (for EC and TDS only) Default value is 1.90%/°C | |||
| TDS Factor | 0.40 to 0.80 (default value is 0.50) | |||
| Probe | HI 76309 4-ring probe, K=1 nominal and built-in temperature sensor (included) | |||
| Power Supply | 4x1.5V AA alkaline type batteries (included), or 12 VDC adapter | |||
| Auto Off | After 5 minutes (can be disabled) | |||
| Casing | IP 67 | |||
| Environment | 32 to 122°F (0 to 50°C); RH 100% | |||
| Dimensions | 7.7 x 3.1 x 2.4" (196 x 80 x 60 mm) | |||
| Weight | 18 oz (500 g) | |||
| (*) Actual conductivity (or TDS) is the non-temperature compensated conductivity (or TDS) value of a solution. | ||||